Vulnerabilities > Medtronic > Mycarelink Monitor 24950 Firmware > Medium
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2019-03-26 | CVE-2019-6540 | Cleartext Transmission of Sensitive Information vulnerability in Medtronic products The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. | 6.5 |