Vulnerabilities > Gehealthcare > Clinical Information Center Mp100D Firmware > Critical

DATE CVE VULNERABILITY TITLE RISK
2020-01-24 CVE-2020-6966 Inadequate Encryption Strength vulnerability in Gehealthcare products
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.
network
low complexity
gehealthcare CWE-326
critical
10.0
2020-01-24 CVE-2020-6962 Improper Input Validation vulnerability in Gehealthcare products
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X CARESCAPE Central Station (CSCS) Versions 2.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, an input validation vulnerability exists in the web-based system configuration utility that could allow an attacker to obtain arbitrary remote code execution.
network
low complexity
gehealthcare CWE-20
critical
10.0
2020-01-24 CVE-2020-6963 Improper Input Validation vulnerability in Gehealthcare products
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
network
low complexity
gehealthcare CWE-20
critical
10.0