Vulnerabilities > Baxter > Em2400 Firmware
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2020-06-29 | CVE-2020-12032 | Cleartext Storage of Sensitive Information vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. | 9.1 |
2020-06-29 | CVE-2020-12024 | Unspecified vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. low complexity baxter | 6.1 |
2020-06-29 | CVE-2020-12020 | Exposure of Resource to Wrong Sphere vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. | 6.1 |
2020-06-29 | CVE-2020-12016 | Use of Hard-coded Credentials vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. | 9.8 |
2020-06-29 | CVE-2020-12012 | Use of Hard-coded Credentials vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. | 6.1 |
2020-06-29 | CVE-2020-12008 | Cleartext Transmission of Sensitive Information vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. | 7.5 |