Vulnerabilities > CVE-2020-12024 - Unspecified vulnerability in Baxter Em1200 Firmware and Em2400 Firmware

047910
CVSS 3.6 - LOW
Attack vector
LOCAL
Attack complexity
LOW
Privileges required
NONE
Confidentiality impact
PARTIAL
Integrity impact
PARTIAL
Availability impact
NONE
local
low complexity
baxter

Summary

Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.