Vulnerabilities > Philips > Intellivue Mx400 Firmware > Medium

DATE CVE VULNERABILITY TITLE RISK
2020-09-11 CVE-2020-16216 Improper Input Validation vulnerability in Philips products
In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart.
low complexity
philips CWE-20
6.5
2020-09-11 CVE-2020-16228 Improper Check for Certificate Revocation vulnerability in Philips products
In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.
low complexity
philips CWE-299
6.4
2018-06-05 CVE-2018-10599 Information Exposure vulnerability in Philips products
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.
high complexity
philips CWE-200
5.3