Vulnerabilities > Philips > Intellispace Cardiovascular > High

DATE CVE VULNERABILITY TITLE RISK
2018-08-22 CVE-2018-14787 Improper Privilege Management vulnerability in Philips Intellispace Cardiovascular and Xcelera
In Philips' IntelliSpace Cardiovascular (ISCV) products (ISCV Version 2.x or prior and Xcelera Version 4.1 or prior), an attacker with escalated privileges could access folders which contain executables where authenticated users have write permissions, and could then execute arbitrary code with local administrative permissions.
local
low complexity
philips CWE-269
7.8
2017-11-17 CVE-2017-14111 Insufficiently Protected Credentials vulnerability in Philips Intellispace Cardiovascular and Xcelera
The workstation logging function in Philips IntelliSpace Cardiovascular (ISCV) 2.3.0 and earlier and Xcelera R4.1L1 and earlier records domain authentication credentials, which if accessed allows an attacker to use credentials to access the application, or other user entitlements.
network
low complexity
philips CWE-522
7.2