Vulnerabilities > Hospira > Critical

DATE CVE VULNERABILITY TITLE RISK
2019-03-26 CVE-2014-5401 Code Injection vulnerability in Hospira Mednet 5.8
Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system.
network
low complexity
hospira CWE-94
critical
10.0
2015-07-06 CVE-2015-3955 Improper Restriction of Operations within the Bounds of a Memory Buffer vulnerability in Hospira Lifecare Pcainfusion Firmware 5.0
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
network
low complexity
hospira CWE-119
critical
10.0
2015-07-06 CVE-2014-5406 Insufficient Verification of Data Authenticity vulnerability in Hospira Lifecare Pcainfusion Firmware 5.0
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.
network
hospira CWE-345
critical
9.3
2015-04-29 CVE-2015-3459 Permissions, Privileges, and Access Controls vulnerability in Hospira products
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
network
low complexity
hospira CWE-264
critical
10.0
2015-04-03 CVE-2014-5405 Information Exposure vulnerability in Hospira Mednet 5.8
Hospira MedNet before 6.1 uses a hardcoded cleartext password to control SQL database authorization, which allows remote authenticated users to bypass intended access restrictions by leveraging knowledge of this password.
network
low complexity
hospira CWE-200
critical
9.0