Vulnerabilities > Hospira > Lifecare Pcainfusion Firmware > Critical

DATE CVE VULNERABILITY TITLE RISK
2015-07-06 CVE-2014-5406 Insufficient Verification of Data Authenticity vulnerability in Hospira Lifecare Pcainfusion Firmware 5.0
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.
network
hospira CWE-345
critical
9.3
2015-07-06 CVE-2015-3955 Improper Restriction of Operations within the Bounds of a Memory Buffer vulnerability in Hospira Lifecare Pcainfusion Firmware 5.0
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
network
low complexity
hospira CWE-119
critical
10.0
2015-04-29 CVE-2015-3459 Permissions, Privileges, and Access Controls vulnerability in Hospira products
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
network
low complexity
hospira CWE-264
critical
10.0