Vulnerabilities > Hospira > Lifecare Pca5 > Critical
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2015-07-06 | CVE-2014-5406 | Insufficient Verification of Data Authenticity vulnerability in Hospira Lifecare Pcainfusion Firmware 5.0 The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. | 9.3 |
2015-07-06 | CVE-2015-3955 | Improper Restriction of Operations within the Bounds of a Memory Buffer vulnerability in Hospira Lifecare Pcainfusion Firmware 5.0 Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors. | 10.0 |
2015-04-29 | CVE-2015-3459 | Permissions, Privileges, and Access Controls vulnerability in Hospira products The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands. | 10.0 |