2021-03-15 | CVE-2020-27290 | Information Exposure vulnerability in Hamilton-Medical Hamilton-T1 Firmware 2.2.3 In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an information disclosure vulnerability in the ventilator allows attackers with physical access to the configuration interface's logs to get valid checksums for tampered configuration files. | 4.3 |
2021-03-15 | CVE-2020-27282 | Missing XML Validation vulnerability in Hamilton-Medical Hamilton-T1 Firmware 2.2.3 In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attackers with physical access to render the device persistently unusable by uploading specially crafted configuration files. | 4.3 |
2021-03-15 | CVE-2020-27278 | Use of Hard-coded Credentials vulnerability in Hamilton-Medical Hamilton-T1 Firmware 2.2.3 In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface. | 5.2 |