Vulnerabilities > Baxter > Sigma Spectrum Infusion System Firmware > 6.05
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2020-06-29 | CVE-2020-12040 | Cleartext Transmission of Sensitive Information vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.0/6.05/8.0 Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. | 9.8 |
2020-06-29 | CVE-2020-12039 | Use of Hard-coded Credentials vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.0/6.05/8.0 Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed. | 2.4 |
2019-03-26 | CVE-2014-5433 | Credentials Management vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.05 An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. | 9.8 |
2019-03-26 | CVE-2014-5432 | Improper Authentication vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.05 Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. | 9.8 |
2019-03-26 | CVE-2014-5431 | Use of Hard-coded Credentials vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.05 Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. | 6.8 |
2019-03-26 | CVE-2014-5434 | Use of Hard-coded Credentials vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.05 Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. | 9.8 |