2020-06-29 | CVE-2020-12047 | Use of Hard-coded Credentials vulnerability in Baxter Sigma Spectrum Infusion System Firmware 8.0 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials. | 7.5 |
2020-06-29 | CVE-2020-12045 | Use of Hard-coded Credentials vulnerability in Baxter Sigma Spectrum Infusion System Firmware 8.0 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials. | 7.5 |
2020-06-29 | CVE-2020-12043 | Operation on a Resource after Expiration or Release vulnerability in Baxter Sigma Spectrum Infusion System Firmware 8.0 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted. | 7.5 |
2020-06-29 | CVE-2020-12041 | Incorrect Permission Assignment for Critical Resource vulnerability in Baxter Sigma Spectrum Infusion System Firmware 8.0 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. | 7.5 |
2019-03-26 | CVE-2014-5433 | Credentials Management vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.05 An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. | 7.5 |
2019-03-26 | CVE-2014-5432 | Improper Authentication vulnerability in Baxter Sigma Spectrum Infusion System Firmware 6.05 Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. | 7.5 |