Vulnerabilities > Baxter > Prismax Firmware

DATE CVE VULNERABILITY TITLE RISK
2020-06-29 CVE-2020-12037 Cleartext Transmission of Sensitive Information vulnerability in Baxter Prismaflex Firmware and Prismax Firmware
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system.
network
low complexity
baxter CWE-319
5.0
2020-06-29 CVE-2020-12036 Cleartext Transmission of Sensitive Information vulnerability in Baxter Prismaflex Firmware and Prismax Firmware
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system.
network
low complexity
baxter CWE-319
5.0
2020-06-29 CVE-2020-12035 Use of Hard-coded Credentials vulnerability in Baxter Prismaflex Firmware and Prismax Firmware
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration.
local
low complexity
baxter CWE-798