Vulnerabilities > Baxter > Prismax Firmware
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2020-06-29 | CVE-2020-12037 | Cleartext Transmission of Sensitive Information vulnerability in Baxter Prismaflex Firmware and Prismax Firmware Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. | 7.5 |
2020-06-29 | CVE-2020-12036 | Cleartext Transmission of Sensitive Information vulnerability in Baxter Prismaflex Firmware and Prismax Firmware Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. | 7.5 |
2020-06-29 | CVE-2020-12035 | Use of Hard-coded Credentials vulnerability in Baxter Prismaflex Firmware and Prismax Firmware Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. | 4.9 |