Vulnerabilities > Baxter > Em2400 Firmware > Critical
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2020-06-29 | CVE-2020-12016 | Use of Hard-coded Credentials vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. | 9.8 |
2020-06-29 | CVE-2020-12032 | Cleartext Storage of Sensitive Information vulnerability in Baxter Em1200 Firmware and Em2400 Firmware Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. | 9.1 |