Vulnerabilities > Abbott > Anthem Firmware
DATE | CVE | VULNERABILITY TITLE | RISK |
---|---|---|---|
2018-04-25 | CVE-2017-12716 | Cleartext Transmission of Sensitive Information vulnerability in Abbott products Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. | 6.5 |
2018-04-25 | CVE-2017-12714 | Improper Restriction of Power Consumption vulnerability in Abbott products Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. | 6.5 |
2018-04-25 | CVE-2017-12712 | Improper Authentication vulnerability in Abbott products The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. | 8.8 |