Vulnerabilities > CVE-2021-42028 - Out-of-bounds Write vulnerability in Siemens-Healthineers Syngo Fastview

047910
CVSS 7.8 - HIGH
Attack vector
LOCAL
Attack complexity
LOW
Privileges required
NONE
Confidentiality impact
HIGH
Integrity impact
HIGH
Availability impact
HIGH
local
low complexity
siemens-healthineers
CWE-787

Summary

A vulnerability has been identified in syngo fastView (All versions). The affected application lacks proper validation of user-supplied data when parsing BMP files. This could result in an out-of-bounds write past the end of an allocated structure. An attacker could leverage this vulnerability to execute code in the context of the current process. (ZDI-CAN-14860)

Vulnerable Configurations

Part Description Count
Application
Siemens-Healthineers
1

Common Weakness Enumeration (CWE)