Vulnerabilities > CVE-2020-15486 - Unspecified vulnerability in Drtrust Electrocardiogram PEN Firmware 2.00.08
Attack vector
ADJACENT_NETWORK Attack complexity
LOW Privileges required
NONE Confidentiality impact
HIGH Integrity impact
NONE Availability impact
NONE low complexity
drtrust
Summary
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.