Vulnerabilities > CVE-2018-8868 - Unspecified vulnerability in Medtronic products

047910
CVSS 6.9 - MEDIUM
Attack vector
LOCAL
Attack complexity
MEDIUM
Privileges required
NONE
Confidentiality impact
COMPLETE
Integrity impact
COMPLETE
Availability impact
COMPLETE
local
medtronic

Summary

Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.