Vulnerabilities > CVE-2017-12722 - Out-of-bounds Read vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1/1.5/1.6

047910
CVSS 5.3 - MEDIUM
Attack vector
NETWORK
Attack complexity
LOW
Privileges required
NONE
Confidentiality impact
NONE
Integrity impact
NONE
Availability impact
LOW
network
low complexity
smiths-medical
CWE-125

Summary

An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.

Common Weakness Enumeration (CWE)

Common Attack Pattern Enumeration and Classification (CAPEC)

  • Overread Buffers
    An adversary attacks a target by providing input that causes an application to read beyond the boundary of a defined buffer. This typically occurs when a value influencing where to start or stop reading is set to reflect positions outside of the valid memory location of the buffer. This type of attack may result in exposure of sensitive information, a system crash, or arbitrary code execution.